What the instrument actually measures
Each normal mode gives an IR band with strength set by its dipole strength \(D_i\), and a VCD band with signed strength set by its rotational strength \(R_i\). The VCD band is several orders of magnitude weaker than the IR band.
Beer–Lambert with molar concentration \(c\) (mol L⁻¹) and path length \(l\) (cm); \(\Delta\varepsilon\) in L mol⁻¹ cm⁻¹.
Last equality holds at the band maximum for an isolated band with common line shape.
A racemate has zero VCD. A sample with enantiomeric excess ee gives \(\mathrm{ee}\times\) the pure-enantiomer spectrum.
Rotational strength: an electric and a magnetic dipole in the same transition
Circular dichroism needs a transition that moves charge both linearly (electric dipole \(\boldsymbol\mu\)) and circularly (magnetic dipole \(\mathbf m\)). Their projection onto each other decides the sign.
Harmonic approximation: two tensors per atom
Also the only tensor IR intensities need.
Electronic (I) and nuclear (J) parts. The derivative with respect to magnetic field B is the MFP ingredient.
The AAT depends on the choice of gauge origin, and so do finite-basis rotational strengths unless care is taken. Using gauge-including atomic orbitals (GIAOs, London orbitals) makes the computed AATs origin-independent; this is the standard DFT implementation since Cheeseman, Frisch, Devlin and Stephens (1996).
From line strengths to spectra
γ is the half-width at half-maximum; 4–10 cm⁻¹ is typical for solution spectra. The top figure uses exactly these formulas.
The coupled-oscillator couplet
Two degenerate oscillators couple into in-phase and out-of-phase combinations. If their transition dipoles are skewed in space, the two combinations acquire rotational strengths of equal magnitude and opposite sign. The result is a bisignate couplet, whose sign reports the helicity of the pair.
\(\mathbf T_{12}\) is the vector from oscillator 1 to 2 and \(V_{12}\) the coupling. Which combination lies higher, and the overall sign, depend on the sign of \(V_{12}\) and on the convention; the dihedral dependence \(\propto \sin\tau\) does not.
Try τ = 0° or 180°: the dipoles become coplanar, the triple product vanishes and so does the couplet. Shrink the splitting below the line width and the two lobes cancel each other, which is why weakly coupled pairs give weak VCD even at favourable geometry.
FT-VCD: polarization modulation on top of an FTIR
A standard FTIR interferogram encodes wavenumber at audio frequencies. The PEM adds a much faster modulation of the polarization state. Lock-in demodulation at the PEM frequency gives an AC interferogram whose Fourier transform is the VCD, and the ordinary interferogram gives the DC (IR) spectrum.
\(\alpha_0\) is the peak retardation of the PEM. Because \(\alpha_0\propto\tilde\nu\) (ignoring dispersion), \(J_1(\alpha_0)\) varies across the spectrum. It is measured with a calibration set (a birefringent plate between polarizers, for example a CdS multiple-wave plate) and divided out.
From a vial to an assignment
Measure IR and VCD
Choose solvent and concentration so the bands of interest have A ≈ 0.2–0.8. Collect enough blocks that the noise spectrum, measured as the half-sum of the enantiomers or repeated scans, is well below the weakest band you intend to interpret.
CDCl₃, CD₂Cl₂, CCl₄, DMSO-d₆4 or 8 cm⁻¹ resolutionSearch conformational space
Use molecular mechanics or semi-empirical methods (CREST/GFN-xTB is widely used) to enumerate conformers. Keep everything within roughly 3–5 kcal mol⁻¹ for DFT refinement.
MMFF / CRESTdedupe by RMSDOptimize and compute frequencies with DFT
Re-optimize each conformer and compute harmonic frequencies, APTs and GIAO-based AATs. Include an implicit solvent model; add explicit solvent molecules when strong hydrogen bonding to the solvent changes the spectrum.
B3LYP / B3PW91TZ basis, e.g. cc-pVTZ or def2-TZVPPCM / SMDD3 dispersionBoltzmann-average
Weight each conformer's line spectrum by its population. Free energies include thermal and entropic terms but also carry larger errors; compare against electronic energies and treat close calls with care.
Scale and broaden
Scale frequencies (harmonic DFT overestimates them; factors of roughly 0.96–0.99 are common for hybrid functionals) and convolve with a Lorentzian of 4–10 cm⁻¹ HWHM.
Compare and quantify confidence
Match IR first to establish band correspondence, then compare VCD signs band by band. Back the visual comparison with a similarity measure (Section 6), and state which bands drive the assignment.
Quantifying the match
f = experiment, g = calculation, σ = scale factor optimized on the IR. S ranges from −1 to +1 for VCD.
Large |ESI| means the data discriminate well between enantiomers. Tools such as CompareVCD turn S and ESI into a confidence level.
If the angle ζ between the electric and magnetic transition moments is close to 90°, a small change in geometry, functional or solvent can flip the sign of R. Such non-robust modes should not carry the assignment.
Starting parameters for a solution measurement
| Parameter | Typical choice | Why |
|---|---|---|
| Sample amount | 5–15 mg | Recoverable; VCD is non-destructive. |
| Concentration | 0.05–0.3 M | Target A ≈ 0.2–0.8 for the bands of interest; watch for aggregation and dimers of acids and alcohols. |
| Cell | BaF₂ or CaF₂, 50–100 µm | Transparent in the fingerprint region. BaF₂ reaches lower wavenumbers than CaF₂ (roughly 850 vs 1100 cm⁻¹ in thick windows). |
| Solvent | CDCl₃, CD₂Cl₂, CCl₄, DMSO-d₆ | Deuteration moves solvent bands out of the way. Protic solvents can change conformer populations. |
| Resolution | 4 or 8 cm⁻¹ | Solution bands are broad; coarser resolution gives better SNR per unit time. |
| Acquisition | 1–8 h (in blocks) | Blocks let you reject drifting segments and estimate noise. |
| Baseline | ½[VCD(+) − VCD(−)], or minus racemate or solvent | Removes instrument artefacts that are the same for both enantiomers. |
| Noise check | ½[VCD(+) + VCD(−)] | Should be flat. Its amplitude is your practical detection limit. |
Where VCD fits among methods for absolute configuration
| Method | Needs | Strengths | Limits |
|---|---|---|---|
| VCD | mg amounts in solution, IR-transparent solvent | Every chiral molecule has IR bands; many bands to compare; conformational information | Weak signal, long runs; flexible molecules need many conformers |
| ECD | UV–vis chromophore | Very sensitive, small samples | Few broad bands; TD-DFT excited states are less reliable |
| ROA | Raman-active sample, often aqueous | Works well in water; biomolecules | Even weaker signal, fluorescence interference, few commercial instruments |
| X-ray (anomalous dispersion) | Single crystal | Direct and definitive when it works | Crystallization; light-atom structures need a good Flack parameter |
| NMR with chiral derivatizing agents | Reactive functional group (e.g. Mosher esters) | Standard lab equipment | Requires derivatization and an empirical conformational model |
Applications
Pharmaceuticals and natural products. Absolute configuration of drug candidates and isolated natural products without crystallization, including oils and amorphous solids.
Conformation and solvation. VCD responds to hydrogen bonding and to solvent–solute interactions, including chirality transfer into achiral solvent molecules.
Peptides and proteins. Amide I and II couplets distinguish α-helix, β-sheet and polyproline II structure; used for fibril and aggregate characterization.
Supramolecular and solid state. Films, mulls and matrix-isolated samples, where anomalously large VCD can signal long-range chiral order. Linear-dichroism artefacts need extra care here.
Milestones
First liquid-phase VCD spectra reported by Holzwarth, Moscowitz and co-workers.
Stephens, Nafie and Keiderling confirm and extend the measurements with dispersive instruments.
Nafie and Diem demonstrate Fourier-transform VCD.
Stephens formulates magnetic field perturbation theory, giving a practical ab initio route to R.
DFT with GIAOs for AATs implemented (Cheeseman, Frisch, Devlin, Stephens), making DFT-VCD routine.
Commercial FT-VCD instruments and standard DFT codes turn VCD into a routine AC method, especially in industry.
Five questions before your group meeting
Where to go next
- L. A. Nafie, Vibrational Optical Activity: Principles and Applications, Wiley, 2011. The standard monograph, covering both theory and instrumentation.
- P. J. Stephens, F. J. Devlin, J. R. Cheeseman, VCD Spectroscopy for Organic Chemists, CRC Press, 2012. A practical guide to DFT-based absolute configuration.
- P. L. Polavarapu, Chiroptical Spectroscopy: Fundamentals and Applications, CRC Press, 2016. Covers VCD, ECD, ROA and ORD side by side.
- P. J. Stephens, "Theory of vibrational circular dichroism", J. Phys. Chem. 89, 748 (1985). The MFP formulation.
- J. R. Cheeseman, M. J. Frisch, F. J. Devlin, P. J. Stephens, Chem. Phys. Lett. 252, 211 (1996). DFT and GIAO implementation of AATs.
- G. Holzwarth, E. C. Hsu, H. S. Mosher, T. R. Faulkner, A. Moscowitz, J. Am. Chem. Soc. 96, 251 (1974). The first VCD measurement.
- V. P. Nicu, E. J. Baerends, Phys. Chem. Chem. Phys. 11, 6107 (2009). Robust normal modes.
- E. Debie, E. De Gussem, R. K. Dukor, W. Herrebout, L. A. Nafie, P. Bultinck, ChemPhysChem 12, 1542 (2011). Confidence level algorithm (CompareVCD).
Check volume and page numbers against the publisher before you cite them in a thesis.